Guidance on medical device establishment licensing (GUI-0016): Other regulatory requirements
This document is for reference only. It will not come into effect until December 1, 2026. For the current version, refer to Guidance on medical device establishment licensing (GUI-0016).
Disclaimer: This document doesn't constitute part of the Food and Drugs Act or its regulations. If there is any inconsistency or conflict between the act or regulations and this document, the act or the regulations take precedence. This document is an administrative document to help the regulated party comply with the act, the regulations, and the applicable administrative policies.
Date issued: October 30, 2026
Effective date: December 1, 2026
Replaces: Guidance on Medical Device Establishment Licensing (GUI-0016) version 10 (December 14, 2024)
On this page
Inspections
Health Canada may inspect anyone who has an MDEL to ensure they comply with the Food and Drugs Act and its Medical Devices Regulations (MDR).
Learn more:
Maintaining records
Distribution records
Manufacturers, importers, and distributors of a medical device must each maintain a distribution record for each device. You must also establish, implement and maintain documented procedures for how you create and keep your distribution records.
Your distribution record must contain enough information so that you can quickly recall any medical device completely from the market. Your procedures should specify how long you retain your distribution records and how you will maintain them.
Section 55 of the MDR specifies that manufacturers, importers and distributors shall retain the distribution record maintained in respect of a medical device for the longer of:
- the projected useful life of the device
- 2 years after the date the device is shipped
Refer to sections 52 to 56 of the MDR for more information on distribution record requirements, including:
- minimum retention periods for distribution records
- how to maintain records to allow for quick retrieval
Complaint handling
Manufacturers, importers and distributors must each maintain records of reported problems concerning the performance or safety for all medical devices they have sold. These records must include all actions taken to respond to these problems. You must also establish, implement and maintain documented procedures for handling complaints.
Learn more:
Reporting problems
Incident reporting
Importers and manufacturers of medical devices must each establish, implement and maintain documented procedures for incident reporting. You must also provide both a preliminary and a final report to Health Canada about:
- devices that they sold in Canada for which incidents that took place inside Canada have been brought to their attention, or
- Class I devices for which incidents that took place inside or outside Canada have been brought to their attention, and:
- the manufacturer has indicated, to a regulatory agency of the country in which the incident occurred, their intention to take corrective action, or
- the regulatory agency has required the manufacturer to take corrective action
To avoid duplicate reporting, the manufacturer of the device may permit the importer to prepare and submit the information on their behalf if both of their information is identical. The manufacturer must inform us, in writing, if such permission has been granted to the importer.
However, a manufacturer may not prepare and submit a report on behalf of an importer.
Learn more:
Serious risk of injury to human health
Importers and manufacturers of Class II, III or IV medical devices must submit information to Health Canada within 72 hours after they learn about a serious risk of injury to human health that's relevant to the safety of the device. This is also called a foreign risk notification.
The manufacturer of a medical device may permit the importer of the device to prepare and submit the information on their behalf if both of their information is identical. The manufacturer must inform us, in writing, if such permission has been granted to the importer.
However, a manufacturer may not prepare and submit a report on behalf of an importer.
Importers of Class II, III or IV medical device must establish, implement and maintain documented procedures for foreign risk notification.
Learn more:
Recalls
Medical device manufacturers, importers and distributors must each establish, implement and maintain documented procedures for recalls.
When they recall a medical device, the manufacturer and importer of that device must each send a report to Health Canada, on or before the recall. This report must outline the information specified in section 64 of the MDR. Manufacturers, importers and distributors should comply with section 58 and conduct a recall when required.
As soon as possible after completing a recall, as per section 65 of the MDR, the manufacturer or importer must report to Health Canada:
- the results of the recall
- the actions taken to prevent a recurrence of the problem
A manufacturer may allow the importer to submit the information and documents relating to the recall on its behalf by notifying Health Canada in writing if the information that must be submitted is identical for both of them.
Learn more:
Reporting shortages and discontinuations
Manufacturers and importers of specified medical devices must report shortages and discontinuations in accordance with sections 62.21 to 62.25 of the MDR. A specified medical device refers to any device included in a category on the List of Medical Devices – Shortages and Discontinuation of Sale.
When a manufacturer or importer becomes aware of a shortage of a specified medical device they must submit a shortage report to Health Canada. Similarly, they must report decisions to discontinue the sale of specified medical devices if the discontinuation would result in a shortage. The reports must be submitted within the specified timelines in the MDR and outline the information specified in subsections 62.23(1) and 62.241(1).
Manufacturers and importers must also update submitted shortage or discontinuation reports when they become aware that the information has changed within the specified timelines in the MDR.
Health Canada recommends that manufacturers and importers develop internal written procedures for reporting medical device shortages and discontinuations.
Email the Medical Device Shortages Unit if you have questions about medical device shortage and discontinuation reporting.
Email: MD.shortages.penurie.de.IM@hc-sc.gc.ca
Learn more: