Guidance on clinical evidence requirements for medical devices: Comparator devices

Date published: November 15, 2022

On this page

About comparator devices

Certain devices require adequate device-specific clinical data. However, Health Canada understands that many medical devices are developed by making rapid and incremental changes to the design. In some cases, the proposed device may be so similar to an existing device that device-specific clinical data are not required.

Where pre-clinical bench testing and side-by-side comparisons can objectively demonstrate that the clinical performance of 2 devices is equivalent, then there is less likely a need for device-specific clinical data.

In addition, Health Canada may have already assessed a comparator device to be safe and effective for the same indications for use. This fact alone may be enough evidence of acceptable performance for the subject device, which means that device-specific clinical data may not be required.

If no similar licensed device can be used as a comparator (does not exist), a manufacturer may use publicly available data sources of comparable devices to assess clinical performance of the subject device. This would avoid the need to generate new clinical data for the device. To do so, manufacturers must:

A selected comparator device should have device-specific clinical data for its intended purpose. Health Canada may or may not accept the comparator device clinical data as stand-alone evidence of the subject device's safety and effectiveness. This will depend on how robust and applicable the clinical evidence is and what uncertainties associated with differences, if any, there are in the clinical performance of the 2 devices.

Manufacturers who rely on comparator device information to support the safety and/or effectiveness of their subject device should provide a thorough comparison between the devices. To demonstrate equivalence and relevance, the following characteristics should be compared:

The manufacturer should demonstrate that the clinical performance of the subject device is well understood and explainable based on the device technology from the comparator. The need for a device-specific clinical investigation may depend on the ability of the existing evidence to:

Some device types will continue to require device-specific clinical data due to their nature or risk profile. These include:

Selecting a device for comparison

It's important to select a relevant device for comparison. Manufacturers should not compare devices that:

Manufacturers should choose a device for comparison that's used as the standard of care for a specific indication. Information in the application of the subject device should demonstrate that the approved comparator device is performing as labelled and that risks are both well characterized and appropriately mitigated. The evidence should be current and relative to current best medical practices. It may also include post-market performance data that demonstrate real-world effectiveness and safety.

Multiple devices may be used for comparison if the subject device shares technological characteristics with more than 1 previously licensed or similar medical device.

Criteria for comparing medical devices

To assess performance adequately, the manufacturer should ensure that the comparator is used in a similar context for a sufficient time in a sufficient number of patients. Health Canada will consider if the experience with the comparator is sufficient in order to determine whether the comparable device is suitable.

Other criteria for ensuring that a comparable device is suitable include:

The manufacturer should also demonstrate that the associated data are applicable by demonstrating that the comparator device has been:

The manufacturer should include the supporting non-clinical information within the technical documentation for the subject device.

Differences between the comparator device and the subject device should not influence the clinical safety and effectiveness of the subject device. The manufacturer should address any differences between the comparator device and the subject device that may affect the ability to extrapolate the clinical evidence to the subject device. An example would be a difference in technical characteristics. In this case, the manufacturer should explain why this difference would not result in different clinical outcomes.

Presentation of device comparisons

A table may be used to present similarities and differences between the subject device and the comparator device. Any clinical evidence related to the previously licensed or similar device that supports the safety and effectiveness of the subject device should be referenced.

We have provided an example summary table:

Characteristics of
subject device X
Characteristics in
comparator device Y
Related clinical evidence
Indications for use/intended use for subject device X Indications for use/intended use for comparator device Y Summary of how the clinical evidence supports the subject device given the indications for use/intended use
Clinical characteristics for subject device X Clinical characteristics for comparator device Y Summary of how the clinical evidence supports the subject device given the clinical characteristics
Physical and performance characteristics for subject device X Performance characteristics for comparator device Y Summary of how the clinical evidence supports the subject device given the identified performance characteristics
Biological characteristics for subject device X Biological characteristics for comparator device Y Summary of how the clinical evidence supports the subject device given the biological characteristics

Manufacturers should:

Page details

Date modified: