Archived lists of medical devices for expanded use in relation to COVID-19 under the Medical Devices Regulations (2023-02-22 to 2023-09-15)
Notice to Reader:
This is an archived list. Please refer to the current version of this list.
This List of Medical Devices for Expanded Use in Relation to COVID-19 (the List) is incorporated by reference in section 68.36 of the Medical Devices Regulations (MDR).
Pursuant to section 68.36 of the MDR, the Minister of Health (Minister) may add a licenced medical device or an authorized Class II, III or IV COVID-19 medical device to column 1 of the List.
The Minister may also add an expanded use to column 2 of the List for the device in column 1.
These may be added because the Minister:
- had reasonable grounds to believe that there is an urgent public health need related to COVID-19 for the expanded use of the device
- had sufficient evidence to support the conclusion that the benefits associated with the expanded use outweigh the risks associated with it, given:
- uncertainties relating to those benefits and risks
- the urgent public health need for the device
- determined that the health or safety of patients, users or other persons will not be unduly affected by the expanded use of the device
Pursuant to section 68.37 of the MDR, the Minister shall publish supplementary information on the expanded use of a device in column 1 of the List.
Supplementary information includes:
- a statement of the expanded use
- a statement of the known and potential benefits and risks
- any supplement to the directions for use, unless a supplement is not required for the device to be used safely and effectively
For more information, please refer to:
Column 1 COVID-19 medical device or other medical device |
Column 2 Expanded use |
Supplementary information |
---|---|---|
ID Now COVID-19 Test Device (IO 320087) | Shelf Life Expansion ID NOW COVID-19 assay performed on the ID NOW Instrument is a rapid molecular in vitro diagnostic test utilizing an isothermal nucleic acid amplification technology intended for the qualitative detection of nucleic acid from the SARS-CoV-2 viral RNA in direct nasal, nasopharyngeal or throat swabs from individuals who are suspected of COVID-19 by their healthcare provider within the first seven days of the onset of symptoms. The shelf life of the device has been authorized to be 24 months, which further expands the shelf life from the previous expanded shelf life authorization on November 10, 2021, 6 additional months. |
24-month shelf life expansion (2022-05-20) Shelf Life Expansion The instructions for use as well as the marked expiry date on the product provided with the ID Now COVID-19 test (IO 320087) from the manufacturer includes a stated expiry/shelf life of 9 months. On August 4, 2020, Health Canada extended the expiry date of the device by 3 months. On November 10, 2021, Health Canada further extended this period by an additional 6 months. On May 20, 2022, Health Canada further extended this period by an additional 6 months, for a total shelf life of 24 months. The total expanded shelf life is 15 months (3 months from the first expanded use authorization, 6 months from the second expanded use authorization, and 6 months from the third expanded use authorization). The anticipated benefit of this expanded use is the continued safe and effective use of testing devices beyond the stated expiry date. Through this shelf life extension, the purchased tests can safely be used for the detection of SARS-CoV-2. The potential risks are that users may be confused on the correct shelf life of the purchased product. To mitigate the risk, Health Canada is posting the extended shelf life notification on the web. In addition, users or distributors that have any questions regarding the shelf life of their product should contact Health Canada's Medical Devices Directorate at hc.meddevices-instrumentsmed.sc@canada.ca. |
ID Now COVID-19 Test Device (IO 320087) | Sample Storage Conditions ID NOW COVID-19 assay performed on the ID NOW Instrument is a rapid molecular in vitro diagnostic test utilizing an isothermal nucleic acid amplification technology intended for the qualitative detection of nucleic acid from the SARS-CoV-2 viral RNA in direct nasal, nasopharyngeal or throat swabs from individuals who are suspected of COVID-19 by their healthcare provider within the first seven days of the onset of symptoms. In addition to the existing authorization for use with swabs that have been stored at room temperature (15 - 30 degrees C) for up to one hour prior to testing, the test is also authorized to be used with the following:
|
Sample Storage Conditions (2021-03-05) The instructions for use provided with the ID Now COVID-19 test (IO 320087) includes instructions that swabs should be placed in a capped, clean, unused tube and stored at room temperature (15 - 30 degrees C) for up to 1 hour prior to testing. This remains the preferred storage conditions for samples to be analysed by the ID Now COVID-19 test. However, if workflow does not permit the use of the sample within this period, Health Canada has also reviewed sufficient evidence supporting the stability of a sample that is stored in a capped, clean, unused tube at 2 - 8 degrees C for no more than 24 hours. If the sample will be shipped during this period, the same storage condition and temperature range needs to be respected. The anticipated potential benefit of this expanded use during the pandemic is greater flexibility of workflow in locations where use of the sample within the preferred one-hour period is not possible. This has the potential to positively impact use of the ID Now COVID-19 test workflows in northern, remote and isolated communities of Canada to which the Abbott ID Now COVID-19 has been deployed. The potential risks are that users may inappropriately store the sample in a container that is not a capped, clean, unused tube, or may not respect the temperature limitations. In such cases, sample degradation could be possible. |
BD Veritor System for Rapid Detection of SARS-CoV-2 (318420) | Visual interpretation of results. BD Veritor System for Rapid Detection of SARS-CoV-2 device is a chromatographic digital immunoassay intended for the direct and qualitative detection of SARS-CoV-2 nucleocapsid antigens in nasal swabs from individuals who are suspected of COVID-19 by their healthcare provider within the first five days of the onset of symptoms. The results are interpreted using the BD Veritor Plus Analyzer. The indication for use has been expanded to include visual interpretation of the results. The expanded use, allowing visual interpretation, does not extend to asymptomatic serial testing. When performing asymptomatic serial testing, the instructions for use should be followed and the BD Plus Analyzer should be used to interpret the results. |
Visual interpretation of results (2021-10-08) The instruction for use should be followed for the sample collection, preparation and loading. Once the sample has been loaded wait the prescribed time of 15 minutes before interrupting the results. When the test is ready, elevate the device, if necessary, to a position where the device reading window is optimally positioned for user visualization. Slowly tilt the device back and forth to remove unnecessary glare. Examine the device reading window for the visual presence of lines in the Control (C), Test (T) and Non-specific (N) regions. Visual interpretation of the results:
The limit of detection when interpreting the results visually increases to 7.8 x 103 TCDI50/ml. This increase in the limit of detection carries a risk. However, should there be doubt in the visual read interpretation, the device can be inserted into the analyzer and the standard steps outlined in their instructions for use should be followed. The risks associated with increased limit of detection should be mitigated by utilizing the analyzer in cases of doubt. The anticipated benefit of this expanded use is the increase flexibility in workflow while using the BD Veritor System for Rapid Detection of SARS-CoV-2 (IO 318420). The safety and effectiveness of this device is maintained while utilizing this expanded use. |
BD Veritor System for Rapid Detection of SARS-CoV-2 (318420) | Shelf life extension BD Veritor System for Rapid Detection of SARS-CoV-2 device is a chromatographic digital immunoassay intended for the direct and qualitative detection of SARS-CoV-2 nucleocapsid antigens in nasal swabs from individuals who are suspected of COVID-19 by their healthcare provider within the first five days of the onset of symptoms, or from individuals without symptoms or other epidemiological reasons to suspect COVID-19 when tested twice over two or three days with at least 24 hours, and no more than 48 hours, between tests. The results are interpreted using the BD Veritor Plus Analyzer. The shelf life of this device has been authorized to be 16 months, an extension of 4 months from the previously authorized 12 months. |
Shelf life extension (2021-10-18) The instructions for use as well as the marked expiry date on the product provided with the BD Veritor System for Rapid Detection of SARS-CoV-2 testing device (IO 318420) from the manufacturer includes a stated expiry/shelf life of 12 months. This expiry date is based on an amendment authorization for IO 318420 extending the shelf life from 6 months to 12 months. Health Canada has extended the expiry date by an additional 4 months to a total of 16-months. Becton Dickson and Company, the manufacturer, will not be incorporating the 16 month shelf life on their packaging, labelling, or instructions for use. Therefore, the stated shelf life/expiry date has 4 additional months before the device expires. The anticipated benefit of this expanded use is the continued safe and effective use of testing devices beyond the stated expiry date. Through this shelf life extension, the purchased tests can safely be used in the detection of SARS-CoV-2. In order to aid in the determination of the correct expiry date for your device Becton Dickson and Company has provided additional instructions based on the current expiry date of the device and the REF number of the device. Follow the guide below to determine the expiry date of your device. For device REF number 256089: If the expiry date printed on the label ends in 2021, Then, the new expiry date = Expiry date printed on the label + 10 months For example, products with a printed expiry date of 2021-06-09, now have a new expiry date of 2022-04-09 If the expiry date printed on the label ends in 2022, Then, the new expiry date = Expiry date printed on the label + 4 months For example, products with a printed expiry date of 2022-01-09, now have a new expiry date of 2022-05-09 For Device REF number 256082: The new expiry date = Expiry date printed on the label + 10 months. For example, products with a printed expiry date of 2022-01-01, now have a new expiry date of 2022-11-01. |
Cepheid Xpertxpress Test Device (IO 312836) |
Xpertxpress SARS-CoV-2 is a rapid molecular in vitro diagnostic test that is performed on the GeneXpert Dx and GeneXpert Infinity systems that currently utilizes nasopharyngeal swab, nasal wash, and aspirate specimens under its existing authorization. Through a clinical trail and a limit of detection study, the performance of the device met Health Canada's requirements for expanded use authorization. In addition to the existing authorized sample types, the gargle and spit sample type is authorized for use with the Xpertxpress SARS-CoV-2 rapid molecular test. In order to collect gargle and spit samples, the instructions provided by the BC Ministry of Health and Centre for Disease Control should be followed, namely:
Further instructions are available on the BC website for children and youth (EN and FR) and adults (EN and FR) as well as video demonstrations that are also available in American Sign Language. Collected samples are stable for up to 5 days at 4 degrees C and up to 48 hours at room temperature. Specimens should be transported refrigerated (ice packs in insulated containers) to avoid exposure to higher temperatures if transportation delays are expected. |
Expanded use (2021-04-15) The samples used to validate the use of gargle and spit samples on the Cepheid XpertXpress SARS-CoV-2 test were collected in Fangye Cryogenic tubes with funnels (manufactured by TaiZhou Huangyan Fangye Technology Development Co. Ltd) using Sterile Saline Addipak®0.9% (5 mL) (NPN 00598127, manufactured by Teleflex Medical, Research Triangle Park, NC, USA). The Starplex Scientific Leakbuster ™ Container (90 mL) (manufactured by Starplex Scientific, Etobicoke, Ontario, Canada) was used to collect the samples in combination with the Sterile Saline Addipak®0.9% (5 mL) for clinical purposes. Both sample collection containers are considered appropriate for use with this type of sample. The anticipated potential benefit of this expanded use during the pandemic is greater flexibility in the sample types that can be used with the Cepheid XpertXpress SARS-CoV-2 test. In addition, the gargle and spit sample collection method is less invasive than other sample types, improving the user experience of the test. Participants in the clinical analysis considered the gargle and spit sample collection to be significantly more acceptable than nasopharyngeal samples collected by healthcare professionals. The potential risks are that users may collect samples inappropriately because the instructions for use related to this sample type are not supplied with the product. Further, information on potential interference by endogenous or exogenous substances has not been characterized for this sample type. Confirmatory testing using alternate samples should be considered when results do not concord with expectations based on clinical assessment or other considerations. Additionally, there is a risk that users will seek clarification or additional information from Cepheid related to these sample types. Cepheid is not in a position to provide information about the collection or analysis of gargle and spit samples by the XpertXpress SARS-CoV-2 test. Such questions should be directed to Health Canada's Medical Devices Directorate at meddevices-instrumentsmed@hc-sc.gc.ca. |
Freestyle Libre Flash Glucose Monitoring System |
The FreeStyle Libre Flash Glucose Monitoring System Reader ("Reader")/FreeStyle LibreLink app ("App") when used with a FreeStyle Libre Flash Glucose Monitoring System Sensor ("Sensor") is indicated for measuring interstitial fluid glucose levels in adults aged 18 years and older (with or without diabetes, including pregnant women) in all hospital and all professional healthcare settings including patients receiving medical intervention/therapy for COVID-19. The Reader/App and Sensor are designed to replace blood glucose testing in the self-management of diabetes, including dosing of insulin. Treatment decisions should not be based on real time Sensor glucose readings alone and instead should consider all the information on the results screen. |
Expanded use (2021-02-23) Additional Information for Pregnant Women: The system is not intended to be used by pregnant women for self-management of diabetes. The system readings can only be used for pregnant women, to replace blood glucose testing, when used by healthcare professionals in hospital and professional healthcare settings. The system can replace blood glucose testing except in a few situations. These are the times when a blood glucose test should be performed before deciding what to do or what treatment decision to make as sensor readings may not accurately reflect blood glucose levels:
Known and Potential Benefits: The anticipated benefit of this expanded use during the pandemic is to allow for timely access to glucose measurements in these patients while minimizing person-to-person contact, reducing the risks of both patient infection and the spread of infection to health care workers, and conserving personal protective equipment. Known and Potential Risks:
*Scott, Eleanor M., et al. Accuracy, User Acceptability, and Safety Evaluation for the FreeStyle Libre Flash Glucose Monitoring System When Used by Pregnant Women with Diabetes. Diabetes Technology & Therapeutics, 20.3 (2018):180-188
Supplemental Labeling: Additional information for Pregnant Women During COVID-19 Pandemic, Shipper Insert, Rev C |
Bravo Autoclave BRAVO 17V and BRAVO 21V (Licence No.: 79049) Device identifiers: |
To provide decontamination solutions for compatible N95 respirators during the SARS-CoV-2 coronavirus pandemic. This expanded use allows for the emergency use of the BRAVO 17V or 21V Porous Wrapped 121°C Cycle in decontaminating compatible N95 flat-fold style respirators (3M 1870+, 9210+ models only), for single‐user reuse by healthcare personnel. |
Expanded use (2021-02-21) The Bravo Autoclaves are not compatible for:
There are no safety or effectiveness issues with regards to the expanded use of this device. The anticipated potential benefits of the expanded use during the pandemic include increased availability of safe and effective respirators for use, reducing respirator supply needs. There are no known additional risks with the expanded use of the device. Additional labelling has been developed by the company, in keeping with other respirator reprocessing devices, which includes the following:
|
Angiographic syringes (Licence No.: 98999) Device identifiers:
|
The syringes used together with hypodermic needles are intended to inject the drug for the patient or take out the blood from the patient. |
Expanded use (2020-12-11) There are no safety or effectiveness issues with regards to the expanded use. The anticipated potential benefits of this expanded use during the pandemic include availability and an additional safe and effective syringe for vaccine use. There are no known additional risks involved with the expanded use of this device. No additional labelling is required for this use. |
Toraymyxin (Pmx-20r) (Licence No.: 63404) |
When TORAYMYXIN is used for COVID-19 patients, indications are as follows:
|
Expanded use (2020-04-20) Limited data from published cases of TORAYMYXIN used for acute respiratory distress due to viral pneumonia and acute exacerbation of interstitial pneumonia suggest that TORAYMYXIN Therapy may:
Additional Warnings and Precautions:
Addendum to Instructions for Use of TORAYMYXIN (PMX cartridge) as a COVID-19 Medical Device:
|
Hemosphere Advanced Monitoring System (Licence No.: 99253) |
The HemoSphere advanced monitoring platform is intended to be used by qualified personnel or trained clinicians in a critical care environment for monitoring of hemodynamic parameters in patients undergoing treatment, including for sepsis, acute respiratory distress syndrome (ARDS), or COVID-19, in a hospital setting. |
Expanded use (2020-04-24) Safety and effectiveness was supported by literature studies, clinical guidelines, and acceptable marketing history in Canada and globally. Additional Warnings and Precautions: N/A Addendum to Instructions for Use: Version EW-CAN-IO-1.0, original 2019-09-10 |
Freestyle Libre Flash Glucose Monitoring System (Licence No.: 99351) |
The FreeStyle Libre Flash Glucose Monitoring System Reader ("Reader")/FreeStyle LibreLink app ("App") when used with a FreeStyle Libre Flash Glucose Monitoring System Sensor ("Sensor") is indicated for measuring interstitial fluid glucose levels in adults aged 18 years and older (with or without diabetes) in all hospital and all professional healthcare settings including patients receiving medical intervention/therapy for COVID-19. In home setting, FreeStyle Libre system is indicated for measuring interstitial fluid glucose levels in adults aged 18 years and older with diabetes. The Reader/App and Sensor are designed to replace blood glucose testing in the self-management of diabetes, including dosing of insulin. Treatment decisions should not be based on realtime Sensor glucose readings alone and instead should consider all the information on the results screen |
Expanded use (2020-04-24) Safety and Effectivenes: Supported by a review of the published literature regarding the use of the system in hospitalized patients (both critically ill and non-critically ill populations), and sensory accuracy performance testing with risk assessment which provided assurance that the benefit of reduced patient contact and reduced usage of personal protective equipment outweighs the risks associated with potential suboptimal sensor accuracy in hospitalized patients. Additional Warnings and Precautions: Hospitals should consider whether they have the resources and expertise necessary to adequately implement Continuous Glucose Monitoring (CGM) system use in their facility and provide appropriate training to healthcare providers. The FreeStyle Libre system is to be used in patients to perform glucose scans by themselves or with the aid of caregivers or healthcare professionals. CGM glucose results are less accurate than blood glucose results obtained using traditional testing methods (e.g., lab glucose, blood glucose meters) in hospital settings. Users should consider all CGM glucose information (e.g., trend) along with individual glucose values, and interpret CGM results in the context of the full clinical picture. CGMs may be subject to interferences that may generate falsely high and falsely low glucose readings. Levels of interference depend on drug concentration; substances that may not significantly interfere in non-hospitalized patients may interfere when used in the hospital setting because of higher dose levels. Most drugs used in hospital or critical care settings have not been evaluated and their interference with CGMs is unknown. Known interferences for FreeStyle Libre system include a positive bias due to high doses of ascorbic acid (Vitamin C) and a slight negative bias when taking salicylic acid (used in some pain relievers such as aspirin). Other reducing drugs / compounds may also interfere. Addendum to Instructions for Use: Additional information for Hospitalized Patients During COVID-19 Pandemic, Shipper Insert, 2020 |
Dexcom G6 Continuous Glucose Monitoring System (Licence No.: 102449) | The Dexcom G6 Continuous Glucose Monitor may be used in hospital to monitor critically ill patients. | Expanded use (2020-04-25) Safety and Effectiveness: It is hoped that the use of real time continuous glucose monitoring devices, such as the Dexcom CGM system, will help reduce personal protective equipment (PPE) usage and hospital staff exposure during the current COVID-19 pandemic. Expanded use was supported by literature on the use of continuous glucose monitoring in hospitalized or critical care settings as well as recent article related to glycemic characteristics and clinical outcomes of COVID-19 patients Hospitalized in the United States. Additional Warnings and Precautions: Continuous glucose readings do have interpersonal variability and have not been tested on COVID-19 related patients. Hospitals should continue to use their existing protocols to manage and treat patient glucose levels. Real time CGM can be used to provide remote monitoring and glucose trends to aid in glucose management and medical decision making. CGM glucose results can be less accurate than blood glucose results obtained using traditional testing methods (e.g., lab glucose, blood glucose meters). Users should consider all CGM glucose information (e.g., trend) along with individual glucose values and interpret CGM results in the context of the full clinical picture. CGMs may be subject to interferences that may generate falsely high and falsely low glucose readings. Levels of interference depend on drug concentration; substances that may not significantly interfere in non-hospitalized patients may interfere when used in the hospital setting because of higher dose levels. Most drugs used in hospital or critical care settings have not been evaluated and their interference with CGMs is unknown. Based on its structure, it is unlikely hydroxychloroquine would interfere with Dexcom CGM. Known interferences vary by CGM brand and can include:
Poor peripheral blood perfusion may cause inaccurate sensor readings. CGM results should be interpreted considering accompanying patient conditions and medications. Other clinical conditions may also cause inaccurate readings. Dexcom G6 CGM System has not been tested in MRI/CT/Diathermy situations. Remove CGM prior to MRI/CT/Diathermy tests. See, Dexcom's "Using Your G6" guide for warnings and precautions. Addendum to Instructions for Use: Fact Sheet for Healthcare Providers: Use of Dexcom Continuous Glucose Monitoring Systems During the COVID-19 Pandemic [Doc. Ref.: LBL018932] |
Dexcom G6 Continuous Glucose Monitoring System (Licence No.: 102449) | The Dexcom G6 Continuous Glucose Monitoring System may be used by pregnant women with diabetes mellitus as an adjunctive device to complement, not replace, information obtained from standard home glucose monitoring devices. Treatment decisions should be based on measurements obtained using a blood glucose meter as per standard of care self-monitoring of blood glucose. | Expanded use (2020-07-10) In the general adult population, the Canadian Medical Device Licence for the Dexcom G6 Sensor labels it to be worn on the abdomen (belly). For pregnant women under this Interim Order Expanded Use, the Dexcom G6 Sensor is being authorized to be worn on the abdomen (belly) or back of the upper arm. Patients are encouraged to speak to their care providers about using the remote glucose data sharing capabilities of the Dexcom G6 System, Dexcom Share App, and Dexcom CLARITY, to share their glucose data with their care providers in an effort to limit the necessity for direct contact between patients and their care providers during the pandemic. The anticipated potential benefits of this expanded use during the pandemic, based on results of a clinical study of the Dexcom G6 System in pregnant women with diabetes mellitus, recommendations of Diabetes Canada and American Diabetes Association guidelines, and related clinical literature for use of continuous glucose monitoring in pregnant women are: reduced contact between patients and their care providers, thus limiting the potential spread of COVID-19, and; to help identify periods of hypoglycemia and hyperglycemia and help achieve glycemic targets. The potential risks, based on the available information, are:
|
Medtronic Enlite Sensor and compatible devices: Enlite Glucose Sensor (Licence No.: 90691) iPro 2 Continuous Glucose Monitoring (CGM) System (Licence No.: 85706) Minilink Real Time Transmitter (Licence No.: 73839) Paradigm Insulin Infusion Pump (Licence No.: 62859) Guardian Link Transmitter (Licence No.: 97803) |
The Medtronic Enlite Sensor and compatible devices: Paradigm Veo (MMT-554/754) insulin pump (with Minilink Real Time Transmitter), Minimed 630G insulin pump (with Guardian Link Transmitter), iPro2 System, may be used in pregnant women with diabetes mellitus. The glucose readings from these devices are intended to be used to complement, not replace, information obtained from standard home glucose monitoring devices. Treatment decisions should be based on measurements obtained using a blood glucose meter as per standard of care self-monitoring of blood glucose. | Expanded use (2020-07-10) Patients are encouraged to speak to their care providers about using the Carelink software to share their glucose data with their care providers in an effort to limit the necessity for direct contact between patients and their care providers during the pandemic. The anticipated potential benefits of this expanded use during the pandemic, based on results of the CONCEPTT randomised controlled trial, recommendations of Diabetes Canada and American Diabetes Association guidelines, and related clinical literature for use of continuous glucose monitoring in pregnant women are: reduced contact between patients and their care providers, thus limiting the potential spread of COVID-19, and; to help identify periods of hypoglycemia and hyperglycemia and help achieve glycemic targets. The potential risks, based on the available information, are:
|
PanbioTM COVID-19 Ag Rapid Test Device (IO 320577) | The Panbio™ COVID-19 Ag Rapid Test Device is a lateral flow immunoassay intended for the qualitative detection of nucleocapsid protein antigen from SARS-CoV-2 in nasopharyngeal swabs from individuals who are suspected of COVID-19 by their healthcare provider. As an alternative to nasopharyngeal swabs, the test is also authorized to be used with the following foam nasal swabs: FOAM SWAB Nasal (COVID-19 AG) (Jiangsu Changfeng Medical Industry Co., Ltd.) FOAM SWAB Nasal (COVID-19 AG) (Noble Biosciences Inc.) (Item code SS-2W) FOAM SWAB Nasal (COVID-19 AG) (MRC) |
Expanded use (2021-02-16) The instructions for use provided with the PanbioTM COVID-19 Ag Rapid Test Device includes instructions for collecting a nasopharyngeal swab only. Users should not follow those sample collection instructions while using a foam nasal swab as this could cause discomfort, pain, or potentially injury. Appropriate nasal swab sample collection instructions and training must be provided to users who will employ the test with authorized nasal swabs instead of nasopharyngeal swabs. The anticipated potential benefit of this expanded use during the pandemic is greater user acceptability of the test when paired with nasal swab collection procedures rather than nasopharyngeal swab collection procedures. The potential risks are that users may inappropriately follow the instructions provided with the test kit for collection of a nasopharyngeal sample while using a nasal swab. The risk associated with this is that the patient may experience discomfort, pain, or potentially injury. This risk must be mitigated by ensuring that anyone who will use this test with nasal swabs is appropriately instructed in nasal swab sample collection technique. Additionally, the test has been validated for use only with the listed foam nasal swabs. The use of other nasal swabs has not been validated with this test at this time. |
Dolphin Neurostim (Licence No.: 67326) |
Instructions for Use for Dolphin Neurostim (NS) applied as a vagal nerve stimulator (VNS) is authorized for emergency use in the acute treatment of adult patients with known or suspected COVID-19. Intended for those:
Using non-invasive VNS applied to the auricular branch of the vagus nerve (ABVN) during the COVID-19 pandemic can help with symptom relief. |
Expanded use (2022-04-01) There are no safety or effectiveness issues with regards to the expanded use. |
Page details
- Date modified: