Guidance on quality requirements for biocides: Stability and shelf life

On this page

Application requirements

You are responsible for ensuring that:

Some biocides have more risk or uncertainty associated with them than others. For these biocides, we require that the initial application include stability data. This data is needed to demonstrate that the biocide's benefits continue to outweigh its risks throughout its shelf life considering related uncertainties.

Biocides that require stability data in their initial applications are those that are:

Conducting stability studies

You can determine the shelf life and expiry date of your biocide by conducting stability studies and considering how your biocide degrades. New chemicals (including impurities) can form as ingredients degrade, which can affect the benefit-risk profile of your biocide.

You can determine the shelf life of your biocide by using either:

You should maintain records of the results and conclusions from stability testing for either:

You should also re-evaluate your biocide's shelf life when there are changes to, for example, the:

If your biocide's composition changes or no longer meets your approved specifications, it could affect your biocide's safety, efficacy and quality. Your biocide should have a shelf life of at least 6 months, as products are generally not expected to be used within 6 months after manufacturing. However, a shorter shelf life may be acceptable if supporting information is provided on its suitability.

Stability testing should show:

Good laboratory practice

When testing for stability, you should use good laboratory practice (GLP). This will help to ensure your data is of high quality and is reliable.

Choose a testing laboratory that complies with our GLP compliance requirements. There are many test laboratories in North America that are able to conduct tests that meet these requirements.

Other types of GLP compliance standards include those published by the:

For non-GLP compliant stability data, you should ensure that your chosen alternate standard are equivalent to GLP certification. F or example, certain International Organization for Standardization (ISO) certification standards are equivalent. This scenario is more common for testing done by European laboratories.

If you propose to use non-GLP compliant data (for example, non-GLP studies or GLP studies with outages), we will assess whether the deviations from GLP or alternative standards are acceptable. You should include rationales to demonstrate:

We will assess whether the deviations from GLP or alternative standards are acceptable. While we will consider all rationales, we cannot guarantee that we will accept the submitted data.

Contact us if you want to verify the acceptability of your non-GLP compliant data prior to submitting your application.

Testing method

Your stability testing methods should include:

Testing of only the smallest package of the same construction and material would be acceptable for most products. If you plan to only test the smallest package size for stability, it is recommended to confirm the acceptability of the stability protocol with us prior to initiating the testing.

There are different test methods for stability testing. You can find details of stability testing options in Table 1. At each test point, you should determine:

Data collection

All stability studies should use at least 1 batch of the same biocide product for each data collection stage.

Table 2: Stability testing for biocides
Type of testing Conditions Duration Frequency Shelf life
Real time (ambient temperature) 20°C or 25°C relative humidity 50% Proposed shelf life 0, 3, 6 and 12 months Equal to the test duration
Real time (warehouse conditions)
Supplementary
Expected warehouse and storage temperature extremes (such as 0°C to 50°C) at least 6 months Every 6 months after for studies that are longer than a year
AcceleratedFootnote 1 54°C 14 days 0 and 14 days 1 year
AcceleratedFootnote 2 for 1 of these conditions and their corresponding duration and frequency 54°C 14 days 0 and 14 days 2 years
50°C 4 weeks 0, 4 weeks
45°C 6 weeks 0, 3, 6 weeks
40°C 4 months 0, 2, 4 months
40°C 6 months 0, 3, 6 months 3 years
35°C 9 months 0, 3, 6, 9 months

Note, if the directions for storage on the label indicate storage at freezing temperatures, additional information may be required to confirm stability using freeze-thaw stability testing.

Footnote 1

This can be considered instead of a study at real time.

Return to footnote 1 referrer

Footnote 2

You should commit to provide real-time stability studies following authorization to supplement accelerated studies.

Return to footnote 2 referrer

Performance standard for significant change

Results of your stability testing should consider any significant changes to your tested batches and to its active ingredients. Significant changes can depend on the nominal concentration of the product:

At the end of the test period, you should also examine for other significant changes, such as:

Significant changes such as these may indicate failed stability, which would have a major impact on the quality of the biocide. Thus, you would need to file a post-authorization application for a major change to change your biocide's shelf life and specifications, as appropriate.

More information can be found in the post-authorization changes for biocides guidance.

Referencing another biocide

You can also determine your biocide's shelf life using information from other sources, when appropriate. You may be able to use this flexibility if you're proposing the same shelf life for an identical biocide (for example, has the same formulation or similar formulation with the only difference being in the fragrance and dyes under 1% weight per weight (w/w), packaging and manufacturing process).

In these situations, your application should include information that shows the proposed biocide is comparable to the referenced biocide, including its stability.

Page details

Date modified: