A Burkholderia stabilis outbreak associated with the use of ultrasound gel

CCDR

Volume 49-7/8, July/August 2023: Enteric Diseases: A Major Health Problem in Canada

Outbreak Report

A Burkholderia stabilis outbreak associated with the use of ultrasound gel in multiple healthcare centres in Montréal, Canada, May–October 2021

Christine Arsenault1,2, Josée Harel3, Florence Doualla-Bell3, Yiorgos Alexandros Cavayas1,4, Xavier Marchand-Sénécal2,5, Charles Frenette6,7, Yves Longtin7,8, Linda Lalande1,9, L Marie-Paule Diby1,9, Nadia Desmarais1,9

Affiliations

1 Hôpital du Sacré-Cœur-de-Montréal, Montréal, QC

2 Département de microbiologie, infectiologie et immunologie, Faculté de médecine, Université de Montréal, Montréal, QC

3 Laboratoire de santé publique du Québec/Institut national de santé publique du Québec, Sainte-Anne-de-Bellevue, QC

4 Département de médecine, Faculté de médecine, Université de Montréal, Montréal, QC

5 Hôpital Maisonneuve-Rosemont, Montréal, QC

6 McGill University Health Center, Montréal, QC

7 Infectious Diseases Division, Faculty of Medicine and Health Sciences, McGill University, Montréal, QC

8 Jewish General Hospital, Montréal, QC

9 Service de prévention et contrôle des infections du CIUSSS du Nord de l’Ile de Montréal, Montréal, QC

Correspondence

christine.arsenault.cnmtl@ssss.gouv.qc.ca

Suggested citation

Arsenault C, Harel J, Doualla-Bell F, Cavayas YA, Marchand-Sénécal X, Frenette C, Longtin Y, Lalande L, Diby LM-P, Desmarais N. A Burkholderia stabilis outbreak associated with the use of ultrasound gel in multiple healthcare centres in Montréal, Canada, May–October 2021. Can Commun Dis Rep 2023;49(7/8):314–9. https://doi.org/10.14745/ccdr.v49i78a03

Keywords: outbreak, Burkholderia stabilis, ultrasound gel, Canada

Abstract

Background: Burkholderia stabilis is a non-fermenting, gram-negative bacteria that has previously been implicated in multiple nosocomial outbreaks through the use of contaminated medical devices and substances. This article reports on an outbreak of B. stabilis infections and colonizations, involving 11 patients from five acute care hospitals in Montréal, Canada.

Methods: One sample was not available for testing, but the remaining 10 isolates (91%) were sent for phylogenetic testing. Medical materials and the patients’ environments were also sampled and cultured. Samples were tested using pulsed field gel electrophoresis and multilocus sequence typing.

Results: The outbreak was found to be associated with the use of intrinsically contaminated non-sterile ultrasound gel. Relatedness of the gel’s and the patients’ B. stabilis strains was demonstrated using gel electrophoresis and multilocus sequence typing analyses. The investigation was concluded with a prompt recall of the product, and the outbreak was declared over by the end of October 2021.

Conclusion: Contaminated non-sterile gel caused infections and pseudo-infections in several patients.

Introduction

On July 25, 2021, an unusually high number of requests for consultations (n=3) by the infectious diseases medical team were placed, seeking advice on Burkholderia stabilis bloodstream infections in the intensive care unit of the Hôpital du Sacré-Coeur-de-Montréal, a 440-bed teaching hospital in Montréal, Canada. It raised concerns about a possible outbreak and led to a formal investigation.

Burkholderia stabilis is a non-fermenting, oxidase-positive gram-negative bacillus, and is a ubiquitous environmental saprophyte. This member of the B. cepacia complex has been associated with nosocomial outbreaks of respiratory infections in patients with cystic fibrosis but can also cause non-respiratory infections in other populations through contamination of various medical devices. Washing gloves Footnote 1, chlorhexidineFootnote 2, alcohol-free mouthwashFootnote 3 and medicationFootnote 4 have all been found to be sources of contamination in previous nosocomial outbreaks of B. cepacia complex. Although non-sterile, multi-use, ultrasound gel is appropriate for use on intact skin and on noncritical devices, it is known to support the growth of pathogenic bacteriaFootnote 5 and has been associated with several outbreaks of B. cepacia complex in different settings Footnote 6Footnote 7Footnote 8.

The epidemiological investigation of the outbreak, the phylogenetic investigation, and the subsequent management are described to prevent further cases.

Method

Outbreak detection

On July 25, 2021, an outbreak of nosocomial bloodstream infections of an unknown source was suspected. An investigation was initiated by the infection control team to identify its source and to prevent exposure of additional patients. First, the laboratory information system was queried for previous positive culture specimens for either B. stabilis or B. cepacia complex. One previous positive blood culture (one or more bottle) for B. stabilis was identified on May 30, 2021, but the isolate had been discarded, in the meantime, as per laboratory protocol. This first case was considered as being part of the outbreak, although its isolate did not contribute to the analysis. Hence, a total of four patients were found to have at least one positive blood culture with B. stabilis over the course of six weeks, of which three isolates were available for further analysis. Three patients had their positive culture sampled 48 hours or more after admission, and one had a positive blood culture sampled on the day of admission. A preliminary case definition was established as a positive blood culture for B. stabilis sampled on the third day after hospital admission or later in the three-month period preceding July 25, 2021. This definition was used to be consistent with the case definition of a nosocomial bloodstream infection by the provincial surveillance programFootnote 9. Symptoms were not required to fit the case definition, and an infection diagnosis was not necessary for inclusion. When no symptom was attributed to the bacteria retrieved in a clinical specimen, it was considered either a contaminant or a colonizer. We defined a contaminant as an organism that is detected by culture but believed to be introduced in the process of sampling the bodily fluid or organ and absent in the fluid or organ itself. A colonizer is a saprophyte organism detected by culture but not causing disease.

A case definition for a possible case included any patient with a culture positive for B. stabilis or B. cepacia complex from any site (other than blood), either nosocomial or community-acquired in the three-month period preceding July 25, 2021.

Investigations

First observations were conducted in the intensive care unit department, where all four cases had been identified. On July 29 and July 30, infection control practitioners audited diverse care techniques provided to patients and related procedures including bathing with single-use gloves, oral hygiene, use of thermometers, central venous catheter manipulations, use of sterile water, handling of multi-use ultrasound gel bottles, and disinfection of noncritical devices.

Subsequently, sampling of clean and sterile material was performed and sent for culture. Indwelling central catheter insertion sites were also swabbed. Sampled material included opened and sealed non-sterile ultrasound gel, sterile ultrasound gel, single-use commercial washing gloves, chlorhexidine wipes, sterile water and mouthwash.

Cultures were incubated on 5% blood sheep agar and MacConkey agar for 48 hours at 37°C in ambient air conditions. Morphologically compatible colonies were submitted for identification using the VITEK MS system using Database v3.1 (bioMérieux, France).

Pulse field gel electrophoresis and multilocus sequence typing analyses

Molecular typing analysis of B. stabilis isolates was done by pulse field gel electrophoresis (PFGE) and multilocus sequence typing (MLST)Footnote 10Footnote 11. Pulse field gel electrophoresis was carried out at the Laboratoire de santé publique du Québec. Sequence typing of B. stabilis isolates was performed by the National Laboratory of Microbiology of Canada according to the protocol and primers specified in a public database of MLST sequence dataFootnote 10Footnote 11Footnote 12.

Interventions

Positive cultures sampled from both opened and sealed ultrasound gel containers originating from the intensive care units were obtained on July 30. The use of all similar products was immediately discontinued at the intensive care units of the Hôpital du Sacré-Coeur-de-Montréal and affiliated hospitals. When additional positive cultures were obtained from ultrasound gel containers from other units, all gel bottles were discarded and replaced by an alternate product on August 2.

Montréal Public Health was notified on August 2 of a suspected contamination of ultrasound gel containers. A notice was sent to physicians and laboratories, and clinical specimens from other hospitals in the Montréal area were sent to the provincial public health laboratory.

Provincial health ministry was notified on August 4 and Health Canada was notified on August 6. A formal complaint was filed to the manufacturer on August 4 and the product was recalled the same day. To identify additional outbreak-related cases in other healthcare institutions in the province of Québec, a microbiology database search was conducted in several hospitals that used the same brand of ultrasound gel. Cultures from any sterile sites found to be positive with B. stabilis and B. cepacia complex were listed. The medical charts of patients with positive cultures were reviewed by local infectious disease consultants to determine whether the positive culture represented a true infection, a contaminant, or a colonization. Patients received care and antimicrobial treatment accordingly.

Results

Over the course of the outbreak, a total of 11 cases of infections and pseudo-infections (detection of a colonizer or contaminant in a specimen sent for culture) were found in five Montréal hospitals, of which 10 isolates were available for analysis; eight specimens were collected between July 16 and August 24 and the last two were collected on September 20 and October 18, 2021 (Figure 1). The 11th isolate had been discarded before the outbreak was declared.

Figure 1: Cases of Burkholderia stabilis infections or pseudo-infections, by week of collection of first positive specimen, May–October 2021Footnote a

Figure 1

Figure 1 - Text description

The graph demonstrates the number of cases of infection and pseudo-infections from Burkholderia stabilis reported by week of collection of the first positive specimen for each case. The first event occurred on May 30th. No other case was identified up to the week of June 19th, when three cases were identified, which was the highest number of cases identified in a single week. Most cases were clustered in July and August, with only one additional case per month in both September and October. The last case was reported on October 18th.


The outbreak was considered over by the end of October 2021 as no further cases were reported and the identified source of the outbreak was no longer in operation.

The case definition used to initiate the investigation proved to be too restrictive, as specimens that were genetically related to the outbreak were sampled from both sterile and non-sterile sites. Consequently, the case definition was reviewed and updated on July 30 to include all cases of infection and pseudo-infections with a genetically related strain of B. stabilis recovered from any type of body specimen.

Of the 33 specimens sampled from medical material and the patients’ environment, six collected from different ultrasound gel bottles were positive for B. stabilis. Five of these bottles were factory sealed prior to sampling and one was already open and in use. B. stabilis was the only bacteria identified in culture. All other samples were negative. Isolates were sent to the public health reference laboratory for further analysis. All patients and gel isolates were clonal after sequencing analysis. All but one isolate was considered definitively related on PFGE analysis, displaying a unique PFGE pattern with restriction enzyme Spel (pulsovar A). One was considered likely related to the outbreak strains, exhibiting a closely related Spel pattern (pulsovar A2). In addition, all isolates shared the same MLST profile and were identified as MLST type ST51, confirming their relatedness.

No death has been attributed to an infection associated with this outbreak. While it is possible that medical care episodes were complicated by a positive blood culture, it was not possible to verify or quantify this impact. Patients presented with wide-ranging clinical profiles. The relevant clinical characteristics are reported in Table 1. Since no surgical site infection was reported, surgeries are not included in these reported data.

Table 1: Characteristics of patients with a positive culture with a clonal stain of Burkholderia stabilis
Hospital Reason for admission Procedures involving ultrasound gel prior to or at time of positive cultures Type of specimens Signification of culture result as per infectious disease consultant
1 Trauma

Central line insertion

FAST ultrasound

Blood cultures Infection
2 Cardiac arrest Peripherally inserted central line, transthoracic echocardiogram Endotracheal secretions Colonization
2 Birth (newborn) External fetal monitoring Umbilical cord blood cultures Contaminant
2 Fall Peripherally inserted central line Blood cultures Infection
3 Trauma

Central line insertion

FAST ultrasound

Blood cultures Infection
3 Neurological condition

Peripherally inserted central line

Venous cavography (using surface ultrasound)

Blood cultures Infection
3 Neurological condition Transthoracic cardiac ultrasound Blood cultures Infection
3 Trauma Transthoracic cardiac ultrasound Blood cultures Infection
4 Orthopedic condition Joint ultrasound Synovial fluid Contaminant
5 Congestive heart failure Central line insertion, mesenteric angiogram and embolization, dialysis catheter insertion Bronchoalveolar lavage Infection
Table 1 abbreviations

Abbreviation: FAST, Focused Assessment with Sonography in Trauma


Discussion

This report documents an outbreak of B. stabilis associated with the use of contaminated non-sterile ultrasound gel. Ten clinical isolates and six isolates from opened and sealed ultrasound gel containers showed relatedness by PFGE and MLST analyses, supporting the hypothesis of ultrasound gel being the cause of the outbreak. Most patients were hospitalized in the intensive care unit, and many had a central venous line in place or were intubated.

A similar investigation reporting 119 cases of B. stabilis infections acquired from ultrasound gel produced by the same manufacturer was performed in the United States during the same period Footnote 13; the results of this investigation are consistent with our findings and support our conclusions.

In this outbreak, intrinsic product contamination occurred at the manufacturing stage, as demonstrated by the presence of bacterial strains in sealed gel bottles. Burkholderia cepacia complex organisms are frequently involved in recalls of non-sterile productsFootnote 14. These bacteria are often resistant to biocides used to prevent bacterial proliferation and can survive for prolonged periods in low-nutrient environments. They are a frequent cause of pharmaceutical compound contamination, which can occur because of contaminated surfaces and materials, but most often through the inclusion of contaminated waterFootnote 14. While non-sterile products are vulnerable to contamination, sterile products are manufactured in bacteria-free environments using sterile materials and are therefore much less likely to result in a contaminated product.

The exact mechanism allowing the non-sterile contaminated gel to lead to bacteremia remains unclear and is likely multifactorial. Visual audits did not reveal noncompliance to central line insertion standardsFootnote 15 or non-critical devices disinfectionFootnote 16, but it was noted that ultrasound gel was sometimes removed swiftly using a dry cloth after a bedside examination. However, these audits are by nature limited to a handful of observations. Although non-sterile ultrasound gel is to be used only on intact skinFootnote 5, similar outbreaks related to contaminated gel have occurredFootnote 6Footnote 7Footnote 8. We hypothesize that contamination of the intact skin of vulnerable patients leads to changes in the skin microbiome and colonization with B. stabilis. Burkholderia stabilis is more likely to be a causative organism if these colonized patients subsequently develop a healthcare-associated infection. This suggests sterile gel should be preferred before an impending invasive procedure, as a simple intervention that should reduce the likelihood of similar events.

Most cases occurred over a short period of time, but two cases phylogenetically related to the outbreak occurred after the month of August, with the latest on October 18. While it was not possible to prove this hypothesis, one likely explanation is that some gel bottles were not discarded immediately after the recall and were still in use at the time of the last case. Implementing a temporary prospective surveillance following a product recall could help address this situation and ensure containment of the outbreak.

One strength of this investigation was the fast identification of the source, leading to a prompt recall of the contaminated product. The causal relationship between the cases and the product is supported by the relatedness of the bacterial strains, demonstrated using multiple validated techniques. While it does not prove that the ultrasound gel caused all infections and pseudo-infections, it would be unlikely to observe such genetic similarity between bacteria retrieved in a product used on patients’ skin and cultures due to chance alone, considering the rarity of this pathogen in the infectious disease practice. Although enough isolates were retrieved to support the association, the earliest case’s isolate was not available to be analyzed. No systematic procedure was in place to refer B. stabilis isolates to the public health laboratory; therefore, our samples do not reflect the entire outbreak’s magnitude. We believed that enough data supported the association between the ultrasound gel and the positive cultures of clinical specimens, so no formal case-control study was performed.

This article describes an outbreak of infections and pseudo-infections with B. stabilis, attributed to intrinsically contaminated ultrasound gel. Non-sterile ultrasound gel is vulnerable to contamination with bacterial pathogens at the time of manufacturing, and from human cross-contamination after introduction into clinical use. Healthcare centres must remain aware of the potential for contamination of these products that could lead to multicenter outbreaks. The universal use of sterile single-use ultrasound gel containers could provide a theoretical advantage, but our study cannot determine whether switching to sterile gels could improve patient outcomes. Still, our study supports the generally accepted notion that single use sterile gels should be preferred over multiuse non-sterile gel in at-risk contexts, such as invasive procedures, procedures that involve sterile equipment and for procedures on mucous membranes or non-intact skinFootnote 5.

Authors’ statement

  • CA — Writing original draft, investigation, writing–review and editing
  • JH — Investigation, writing–review and editing
  • FDB — Investigation, writing–review and editing
  • YAC — Writing–review and editing
  • XMS — Investigation, writing–review and editing
  • CF — Investigation, writing–review and editing
  • YL — Investigation, writing–review and editing
  • LL — Investigation, writing–review and editing
  • LMPD — Investigation, writing–review and editing
  • ND — Investigation, writing–review and editing

The content and view expressed in this article are those of the authors and do not necessarily reflect those of the Government of Canada.

Competing interests

YL has received research support from Syneos Health for work unrelated to the current study.

Acknowledgments

We are grateful to the Special Bacteriology Unit of the National Laboratory of Microbiology of Canada for providing Burkholderia MLST speciation and typing. We also gratefully acknowledge the excellent technical assistance of the laboratory employees and infection prevention practitioners of all the hospitals involved.

Funding

None.

References

Footnote 1

Sommerstein R, Führer U, Lo Priore E, Casanova C, Meinel DM, Seth-Smith HM, Kronenberg A, Koch D, Senn L, Widmer AF, Egli A, Marschall J; Anresis; Swissnoso. Burkholderia stabilis outbreak associated with contaminated commercially-available washing gloves, Switzerland, May 2015 to August 2016. Euro Surveill 2017;22(49):17–00213. https://doi.org/10.2807/1560-7917.ES.2017.22.49.17-00213

Return to footnote 1 referrer

Footnote 2

Ko S, An HS, Bang JH, Park SW. An outbreak of Burkholderia cepacia complex pseudobacteremia associated with intrinsically contaminated commercial 0.5% chlorhexidine solution. Am J Infect Control 2015;43(3):266–8. https://doi.org/10.1016/j.ajic.2014.11.010

Return to footnote 2 referrer

Footnote 3

Martin M, Winterfeld I, Kramme E, Ewert I, Sedemund-Adib B, Mattner F. [Outbreak of Burkholderia cepacia complex caused by contaminated alcohol-free mouthwash]. Anaesthesist 2012;61(1):25–9. https://doi.org/10.1007/s00101-011-1954-4

Return to footnote 3 referrer

Footnote 4

Souza Dias MB, Cavassin LG, Stempliuk V, Xavier LS, Lobo RD, Sampaio JL, Pignatari AC, Borrasca VL, Bierrenbach AL, Toscano CM. Multi-institutional outbreak of Burkholderia cepacia complex associated with contaminated mannitol solution prepared in compounding pharmacy. Am J Infect Control 2013;41(11):1038–42. https://doi.org/10.1016/j.ajic.2013.01.033

Return to footnote 4 referrer

Footnote 5

Infection Prevention and Control Canada. Position Statement, Medical Gels. Winnipeg, MB: IPAC; May 13, 2021. https://ipac-canada.org/photos/custom/Members/pdf/21May_Medical_Gels_Position%20Statement_Final.pdf

Return to footnote 5 referrer

Footnote 6

Abdelfattah R, Al-Jumaah S, Al-Qahtani A, Al-Thawadi S, Barron I, Al-Mofada S. Outbreak of Burkholderia cepacia bacteraemia in a tertiary care centre due to contaminated ultrasound probe gel. J Hosp Infect 2018;98(3):289–94. https://doi.org/10.1016/j.jhin.2017.09.010

Return to footnote 6 referrer

Footnote 7

Yamunadevi VR, Ramasubramanian V, Senthur Nambi P, Samundeewari P, Ramakrishnan N. Outbreak of Burkholderia cepacia bacteraemia in a tertiary care centre due to contaminated ultrasound probe gel. J Hosp Infect 2018;100(4):e257–8. https://doi.org/10.1016/j.jhin.2018.04.014

Return to footnote 7 referrer

Footnote 8

Jacobson M, Wray R, Kovach D, Henry D, Speert D, Matlow A. Sustained endemicity of Burkholderia cepacia complex in a pediatric institution, associated with contaminated ultrasound gel. Infect Control Hosp Epidemiol 2006;27(4):362–6. https://doi.org/10.1086/503343

Return to footnote 8 referrer

Footnote 9

National Institute of Public Health of Quebec. Surveillance of pan-hospital nosocomial bacteraemia (BACTOT). Quebec: INSPQ; July 2022. https://www.inspq.qc.ca/infections-nosocomiales/spin/bactot

Return to footnote 9 referrer

Footnote 10

Spilker T, Baldwin A, Bumford A, Dowson CG, Mahenthiralingam E, LiPuma JJ. Expanded multilocus sequence typing for burkholderia species. J Clin Microbiol 2009;47(8):2607–10. https://doi.org/10.1128/JCM.00770-09

Return to footnote 10 referrer

Footnote 11

Jolley KA, Chan MS, Maiden MC. mlstdbNet - distributed multi-locus sequence typing (MLST) databases. BMC Bioinformatics 2004;5(1):86. https://doi.org/10.1186/1471-2105-5-86

Return to footnote 11 referrer

Footnote 12

PubMLST. Burkholderia cepacia complex. [Accessed 2022 Oct 6]. https://pubmlst.org/organisms/burkholderia-cepacia-complex

Return to footnote 12 referrer

Footnote 13

Hudson MJ, Park SC, Mathers A, Parikh H, Glowicz J, Dar D, Nabili M, LiPuma JJ, Bumford A, Pettengill MA, Sterner Jr MR, Paoline J, Tressler S, Peritz T, Gould J, Hutter SR, Moulton-Meissner H, Perkins KM. Outbreak of Burkholderia stabilis Infections Associated with Contaminated Nonsterile, Multiuse Ultrasound Gel - 10 States, May-September 2021. MMWR Morb Mortal Wkly Rep 2022;71(48):1517–21. https://doi.org/10.15585/mmwr.mm7148a3

Return to footnote 13 referrer

Footnote 14

Tavares M, Kozak M, Balola A, Sá-Correia I. Burkholderia cepacia Complex Bacteria: a Feared Contamination Risk in Water-Based Pharmaceutical Products. Clin Microbiol Rev 2020;33(3):e00139–19. https://doi.org/10.1128/CMR.00139-19

Return to footnote 14 referrer

Footnote 15

O’Grady NP, Alexander M, Burns LA, Dellinger EP, Garland J, Heard SO, Lipsett PA, Masur H, Mermel LA, Pearson ML, Raad II, Randolph AG, Rupp ME, Saint S; Healthcare Infection Control Practices Advisory Committee (HICPAC). Guidelines for the prevention of intravascular catheter-related infections. Clin Infect Dis 2011;52(9):e162–93. https://doi.org/10.1093/cid/cir257

Return to footnote 15 referrer

Footnote 16

Centers for Disease Control and Prevention. Rutala WA, Weber DJ, and the Healthcare Infection Control Practices Advisory Committee. Guideline for Disinfection and Sterilization in Healthcare Facilities, 2008. Atlanta, GA; CDC; 2008. https://www.cdc.gov/infectioncontrol/guidelines/disinfection/

Return to footnote 16 referrer

Page details

Date modified: