Guide to validation of terminal sterilization process of drugs (GUI-0074): Process validation, sterilization by ethylene oxide

On this page

Introduction

Ethylene oxide (EO) is mainly used to sterilize items that are sensitive to moist heat or radiation. EO is a toxic, flammable and explosive substance, listed in a schedule 1 of the Canadian Environmental Protection Act.

For guidelines on EO emission, consult:

Factors that influence EO sterilization include:

Product/material definition

You should design the:

For example, you should avoid:

You should also consider the tolerance of the product/material for the required temperatures.

Note: EO must not affect product/material integrity (for example, by causing cracking, phase separation and bio-compatibility).

Sterilizing agent specifications

You should:

Note: You may use EO as a pure gas (100%) or in a mixture of gases (such as carbon dioxide or nitrogen).

EO sterilization parameters

Monitor the following parameters:

Consider the following when designing your EO sterilization cycles:

Include at least the following in your EO sterilization specification:

Note:

Also note:

Humidity used to precondition and condition the product/material should be generated by steam.

Design of EO sterilization cycles

You should define the process to support the validity of process parameters and their tolerances (as defined in your sterilization process specification).

There are 3 basic approaches for validating an EO sterilization cycle:

1. Overkill ½ cycle:

The cycle is developed by performing 3 consecutive studies resulting in total inactivation of the biological indicators to confirm the minimum exposure time. The specified exposure time for the sterilization process should be at least double this minimum time. A fractional cycle of short duration from which BI survivors can be recovered should also be run to demonstrate the recovery technique for BIs exposed to EO gas is adequate.

Overkill cycle calculation:

The cycle is developed by performing time-graded exposures to EO or population-graded BIs exposed to EO, with all other parameters staying the same. The sublethal exposures are then used to calculate the cycle to ensure a 12 log reduction. Following that, routine bioburden monitoring should be performed.

2. Biological indicator (BI)/bioburden cycle:

The cycle is used when the bioburden of the product/material before EO treatment is fairly consistent over time and less resistant than the biological indicator. This sterilization process involves using a microbial challenge population lower than 106 (but not less than 103). Bacillus atrophaeus is commonly used for EO sterilization because of its high resistance. The BI should be distributed throughout the product load and in the same orientation. Placement should include spots that present the greatest challenge to the sterilization cycle.

Note: When developing the EO sterilization cycle and validation studies, you should test biological indicators as soon as possible after exposure to the sterilization cycle. Microbial inactivation continues after the sterilization cycle has been completed due to the presence of EO residues.

3. Absolute bioburden cycle:

This cycle is used when the product bioburden resistance to the EO process is very high (product's bioburden is more resistant than the BI). This can be caused by a number of factors, such as the configuration of the product/material, the quantity or location of the microorganisms, or the bioburden's intrinsic resistance.

The cycle can also be used when the bioburden is very low in resistance and relatively consistent, allowing an optimized sterilization cycle. The absolute bioburden method requires extensive controls of the manufacturing environment in addition to routine product bioburden monitoring and resistance studies.

Cycle development includes:

Representative product is used in EO cycle development studies. An inactivation curve is established for the product bioburden to project the exposure time required to achieve the desired SAL.

For additional guidance on EO sterilization, consult:

Equipment qualification

Your specification for EO equipment should include equipment from 3 phases of the sterilization process:

For your EO equipment specification, include, at a minimum:

Examples of instruments that need calibration:

Page details

Date modified: