Draft guidance on how to interpret ‘significant change’ of a medical device: Types of changes

On this page

Changes to manufacturing processes, facilities or equipment

For the purposes of this section, a manufacturing change impacts how a device is made, but does not intentionally change the device or its packaging and labelling material.

Examples of a manufacturing change would be:

A change to the manufacturing process, facility or equipment can impact the safety or effectiveness of a device. You should consider the impact of all manufacturing changes on the device specifications, performance and material properties. Evaluate these changes using the other applicable sections of this guidance document.

If a change to the manufacturing process, facility or equipment involves changes to the sterilization process, refer to the section on changes to sterilization.

Tolerances that have been increased to allow for more variation in the end product is likely a significant change, as it may impact device performance.

Changes that are likely non-significant are as follows:

For more information, refer to the following sections:

For combination products, there are increased sensitivities associated with the production and application of active pharmaceutical ingredients (APIs).  As changes to the manufacturing process of a combination product that involve APIs are often significant, we recommend that you contact the Medical Devices Directorate by email at meddevices-instrumentsmed@hc-sc.gc.ca for assistance.

Table of examples

Device Proposed change Significant or not significant

Devices sold as non-sterile

Packaging changed from 1 variant of polyethylene to another due to supplier rationalization or cost-saving measures. Verification and stability testing shows integrity has not been compromised.

Not significant

Drug eluting stent

A change to a manufacturing site where a polymer or drug coating is applied to the device.

Significant

Catheters

Change in supplier that extrudes the polymer tubing with no change in finished product performance specifications.

Not significant

Implantable 3D printed devices

Changes to additive manufacturing processes including equipment, raw materials and post-printing processes of the device.

For details, refer to the guidance document Supporting evidence for implantable medical devices manufactured by 3D Printing.

Significant

All devices

A change in material supplier but the design and/or performance specifications of the finished device are not impacted.

Not significant

Biological devices

Change in animal or human tissue supplier.

Note: A change in supplier could impact biological safety requirements for animal or human-derived tissue materials. For more guidance, refer to the section Changes in materials for non in vitro diagnostic devices.

Significant

Changes to manufacturing quality control procedures

Changing or adding new testing or inspection activities for incoming materials, in-process materials and final products is not a significant change if the specifications or acceptance criteria have not been loosened and acceptance quality limits (AQL) have not increased. On the other hand, removing or modifying these activities where the result is that specifications or criteria are loosened or the AQLs are increased is a significant change.

Changes to quality control inspections or tests used to control the quality, purity or sterility of materials and devices that affect the specifications of the finished device or its components, including its packaging, are often considered significant. However, if adding or modifying the inspection or test method provides equivalent or improved assurance of reliability, then the modification is not significant. It must be reported at annual renewal, with applicable justification made to support this modification.   

Examples include changes to the:

For example, changes to the manufacturer's requirements for material acceptance criteria can be considered a significant change if these changes alter the design specifications of the device.

Table of examples

Device Proposed change Significant or not significant

All devices

Modification or removal of a test that characterizes the final product or its performance (for example, bending test for a metal component)

Significant

All devices

Change in design specifications of a device to accommodate a loosening of acceptance criteria as a result of a change in supplied parts or materials.

Significant

All devices

In-process inspection of tubing length in a catheter removed and a final 100% inspection of tubing length introduced later in the manufacturing process.

Not significant

Changes in design

Changes in design range from minor design specification or engineering changes to major changes in operating principles. All design changes must be evaluated, verified and validated according to the accepted procedures recorded in the quality management system.

Significant design changes can include modifications to the:

Control mechanism

A control mechanism is the means by which the action of a device is directed or the output of a device is regulated. As control mechanisms can be complex, any changes to them could alter how the device works. This can impact its risk profile, as well as its safety and effectiveness and therefore changes to the control mechanisms are significant.

Examples of changes:

Operating principles

Operating principles are the means by which a device produces or brings about an intended or appropriate effect. Changes to operating principles are considered significant.

Examples of operating principle changes include switching:

Examples of operating principle changes for an in vitro diagnostic device (IVDD) include:

For more information on operating principle changes that include software, refer to the section on changes to software.

Design specifications

Changes to design specifications can involve:

For an IVDD, design specifications also include performance characteristics (for example, precision, accuracy, sensitivity, specificity, stability).

A change to design specifications may be significant if it influences 1 of the following 3 factors:

  1. the indications for use/intended use
  2. the requirements to have clinical data
  3. the risk profile of the device

1) Indications for use/intended use

Changes to design that affect the indications for use or intended use are significant. They can impact the target patient population (for example, elderly or pediatric patients), the target anatomical areas or the target disease states.

Example of a significant change:

2) Clinical data

Changes to the design specifications that require new clinical data to validate the safety and/or effectiveness of the device are significant.

Examples of a significant change:

3) Risk profile

You should conduct a risk assessment when making a design change. The risk assessment will identify any new risks. If the risk assessment finds that the change would alter the safety and effectiveness of the device, either positively or negatively, the change is significant.

Examples of a significant change:

The change is likely not significant if:

Components or accessories, including Class II

Adding or changing device components or accessories used with the authorized device is a significant change if it:

Modifying an existing or adding a new component or accessory that is intended to be, or labelled to be, compatible with a class III or IV device is a significant change if:

This is true regardless of the classification of the component or accessory, or whether all parts are on the same authorization or authorized separately.

An application to amend the associated Class III or IV device(s) would be required. You would need to:

For more information, refer to the section on Class III and IV amendments for compatible Class II devices.

Human factors of the patient or user interface

A design change that affects how the device is used or how it interacts or interfaces with the patient or user is likely significant.

Examples are changes to:

Cosmetic changes that:

Table of examples

Device Proposed change Significant or not significant

All devices

New foot switch (where there was not one before) added to an electrosurgical generator or other device, and corresponding software added to the operating console.

Significant

Non-active surgically invasive devices

Change in the design characteristics allows for additional or broader indications for use (for example, a smaller-sized hip prosthesis or fracture fixation screw that differs significantly from the previous designs).

Significant

Catheters

Change to the cable design and grip of a steerable ablation catheter, which results in improved deliverability and improved procedural times.

Significant

Catheters

Change to the grip of a steerable ablation catheter to improve comfort for the healthcare professional or to improve the appearance of the device without changing its functionality or any critical forces that could be applied and felt (tactile) by the user.

Not significant

Endocardial lead

Additional polymer support clip added to prevent the electrical connection from becoming dislodged and to increase the axial retention forces.

Significant

Ultrasound transducer

Design of the grip portion updated to improve user comfort. Change does not affect safety or performance.

Not significant

Hemofiltration system, including software controls

New component, combined filter and disposable cartridge added.

Significant

Transurethral thermal system for treating benign prostatic hyperplasia

Software changed to provide automatic control of ramping power, respond to elevated rectal temperatures automatically and adjust power.

Significant

Metallic biliary stent for treating malignant strictures

Two new stent lengths added that are outside the range of previously authorized stent lengths on the same authorization.

Significant

Metallic biliary stent for treating malignant strictures

Two new stent lengths added that are between the previously authorized stent lengths on the same authorization, with no other differences.

Not significant

Total knee system

Longer femoral augments added.

Significant

Total hip system

New bearing surface added.

Significant

Acetabular cups

Change in design to give additional flexibility to implanting surgeons. More holes added to the cups.

Significant

Bone void fillers and putty

Amount of cancellous bone material in the filler increased.

Significant

Anaesthesia machine

Change in the sensor controlling the fresh air proportions.

Significant

Automatic implanted cardiac defibrillator

Internal components, including the capacitors, telemetry coils, batteries and transformers, altered to improve how the device operates.

Significant

Cardiac pacing leads

Two or more electrodes, or a new anchoring mechanism that can result in new indications for use, added and performance claims enhanced.

Significant

Pacing lead

Size of the wire diameter reduced to reduce the overall lead diameter to facilitate introduction into the vessel.

Significant

Patent foramen ovale (PFO) closure device

An 18-millimetre (mm) PFO closure device added to an authorization that includes a 16-mm and a 20-mm PFO closure device. Basic design and delivery system are the same.

Not significant

IVDD test kit

Sample matrix for an IVDD test kit changed from a venous blood sample to a dried blood spot.

Significant

Clinical chemical analyzer

Throughput changed.

Significant

Clinical chemical analyzer

Test volume changed.

Significant

Clinical chemical analyzer

Full automation changed.

Significant

Blood glucose monitor

A new control added.

Significant

Blood glucose monitor

Sample volume reduced by changing the electrode layout, which reduces the test strip sample chamber volume.

Significant

Blood glucose monitor

Alternate sampling location added (for example, abdomen).

Significant

Automated ELISA analyzer

New analyte added (for example, Hepatitis B surface antigen).

Significant

IVDD test kit

The blood collection method for a point-of-care IVDD changes from a capillary draw to a mechanical draw pipette.

Significant

Changes to sterilization

Changes that could affect the effectiveness of the sterilization process or the safety of a sterile device are significant.

The nature of sterilization is such that it is impossible to determine by inspection and testing if the device has been sterilized successfully. Medical devices are considered sterile if you can demonstrate a sterility assurance level (SAL) of 10-6 or better. The sterilization process must be verified and validated, and its performance routinely monitored.

Examples of significant sterilization changes are those where the:

In general, a change that triggers a new or increased risk or produces unexpected results from routine verification and validation is a significant change. Adding a new test acceptance criteria or test method to the existing process to provide the same or better assurance of sterility, reliability or similar safety aspects is a non-significant change.

Non-significant changes would be those limited to:

Examples:

Using a sterilization method that does not adhere to recognized standards may have a significant impact. Consult:

For advice on how to assess the significance of any changes you wish to make to novel sterilization methods, contact the Medical Devices Directorate at meddevices-instrumentsmed@hc-sc.gc.ca.

Note: A change in facility that introduces a new, more challenging organism or bioburden level to the sterilization process which has not been validated before would be significant. This is the case even if the cycle parameters are the same.

Changes to sterile barrier packaging

Packaging changes should also be assessed for their impact on sterilization. In general, any change in packaging characteristics (such as in materials, size, shape, seal width) or configuration (such as in the outer packaging, loading density) is likely significant if it could affect the:

Issues of compatibility between the packaging material and the sterilization process must also be taken into consideration.

Changes to previously approved cycles or configurations

A change to the sterilization method or packaging of a sterile medical device is  likely not significant if:

Table of examples

Device Proposed change Significant or not significant

Sterile medical devices

Sterilization method changed from ethylene oxide (EO) to gamma radiation.

Significant

Sterile medical devices

Biological indicator changed to parametric release.

Significant

Sterile medical devices

Contract sterilizers changed (with no change to cycle parameters) but the method of validating the process is the same.

Not significant

Sterile medical devices

Changes that reduce the SAL.

Significant

Sterile medical devices

Pre-blended EO sterilant changed to EO post-blended with nitrogen where the ultimate concentration of EO in the sterilizer is the same in both cycles, with no change to the critical cycle parameters (delivered dose of EO).

Not significant  

Sterile medical devices

Change from using air (mixture of 80% nitrogen and 20% oxygen) to pure nitrogen in the aeration process to avoid explosive gas mixtures, with no change to the critical cycle parameters (delivered dose of EO).

Not significant  

Sterile medical devices

Air-flow or heating, ventilating and air conditioning (HVAC) system in the manufacturing environment changed, with the sterilization facility physically and environmentally segregated from the manufacturing line and the device bioburden is not increased above previous level.

Not significant  

Sterile medical devices

New alternate EO sterilization facility added with a proposed cycle that is identical to the currently authorized cycle in all critical parameters, but some parameters such as relative humidity are adjusted due to local elevation differences. The new cycle was successfully validated using the overkill method outlined in ISO 11135.

Not significant

Sterile medical devices

Irradiation dose auditing method changed from VDmax to method 1A/1B, method 2 (per ISO 11137) or vice versa

Significant

Sterile medical devices with single pouch packaging

Packaging of a sterile device changed from a single pouch to a new double pouch.

Significant

Sterile medical devices with double pouch packaging

Double sterile barrier packaging replaced with single sterile barrier packaging.

Significant 

Sterile medical devices

A proposed packaging change previously reviewed and approved for a similar device sterilized using the same cycle, where the subject device presents no greater challenge to sterilization than the comparable device and the package materials are identical.

Not significant

Sterile medical devices

Packaging dimensions changed or protective enclosure added within the layers of sterile packaging (for example, resin).

Significant

Sterile medical devices

Concentration or exposure time of EO reduced and successfully validated using a method defined in ISO 11135.

Significant

Changes to software

Software can be embedded within a medical device or an accessory to a medical device or exist as software as a medical device (SaMD). Software is updated easily and frequently, and thus may undergo several changes during its lifecycle. Health Canada considers some of these changes to be significant.

Examples of significant software changes are those that:

Examples of non-significant changes if none of the previous changes apply are those that:

As these are only examples, we recommend that you evaluate additional software changes that can, if not controlled properly, create unexpected software behaviour. Whether a change is significant or not depends on their potential impact on safety or effectiveness.

Other examples of additional software changes that may or may not be significant:

Medical devices that use machine learning (ML), in part or in whole, to achieve their intended medical purpose are known as machine learning-enabled medical devices (MLMD). A predetermined change control plan (PCCP) provides a mechanism to address cases where pre-authorization of planned significant changes is needed to address a known risk. This approach can be beneficial in managing certain known risks with MLMD, such as ML system performance degradation over time due to ML model drift.

For medical devices that have been authorized with a PCCP, subsequent changes made according to the authorized PCCP are not significant. Thus, they do not require the submission of a licence amendment application.

For amendments to a device that are outside of an authorized PCCP, including changes to the PCCP itself, consult:

For more information on submission requirements for MLMDs and PCCPs, consult:

Common software change types and unintended consequences

Software changes can come in several different forms, such as:

Cosmetic changes are often non-significant. However, the changes listed here may be deemed significant, as there may be unintended consequences when executing software code in often complex software environments. For example, an operating system (OS) upgrade may:

You should consider the consequences of the changes to assess whether the change is significant.

A software change is likely not significant if:

Table of examples

Device Proposed change Significant or not significant

Software – skin cancer detection and characterization

A mobile medical app intended for skin cancer detection and characterization is enhanced by updating its model parameters. There are plans to distribute the upgraded app automatically to existing users. The change affects the sensitivity and specificity of the detection algorithm.

The modification has direct impact on the clinical functionality of the device.

Significant

Software – IVD device

A software modification allows an IVD device to improve sample throughput. Modifications include changes to decrease assay times by allowing for shorter sample reaction times. The shorter incubation time may affect diagnostic performance.

Significant

Software – chest X-ray application

A software application using AI-based image analysis of chest X-rays adds a model for detecting endotracheal tubes. The new diagnostic feature introduces a new risk.

Significant

Software – diagnostic ultrasound

A diagnostic ultrasound system has multiple available measurement parameters. Based on a marketing survey of current customers, 1 measurement parameter not directly related to the intended use is removed in a software update. Its removal does not introduce a new risk or modify an existing risk.

Not significant

Software – infusion pump

An infusion pump has 2 occlusion detection alarms: occlusion downstream and occlusion upstream. The software is modified to allow the user to optionally disable 1 alarm. This means the user now has the option to use 1 or both occlusion alarms.  This is a change to existing risk controls in the device.

Significant

Software – IVD analyzer

Records in an IVD analyzer are written into a database table. Certain conditions can cause new data to be merged with existing data in the table, which can lead to an incorrect result. The bug is caused by a misworded software requirement leading to a code error. The requirement was rewritten, code was modified and a new database was created for administrative records to prevent merging. The code was rewritten and a new database was added, causing a change in software specifications.

Significant

Software – operating system

An ultrasound system’s operating system has been upgraded to Windows 10 from Windows 7. No changes are made to any of the available software packages, no code changes were made, and both risk analyses and verification testing showed no unexpected results. The operating system belongs to the same OS family and no code changes have been made to make the software compatible with the new OS. Risk analyses and verification testing showed no unexpected results.

Not significant

Software – operating system

A CT scanner’s operating system has been upgraded to Windows 10 from Windows 7. Small infrastructure and architectural changes were made to the software to allow for compatibility.  The new OS belongs to the same OS family. However, the change could have unintended consequences that could affect the device’s performance.

Significant

Software – operating system

A blood glucose monitor has been modified to make Android and iOS devices compatible (including control of the device). Previously it had been compatible only with Windows PCs. While the underlying functionality of the device has not changed, a complete rewrite of the algorithms for communication with a new family of operating systems requires full testing and risk analysis. Also, risk controls may be very different in a mobile operating system.

Significant

Software – cybersecurity

A security vulnerability is found in a device’s software through routine cybersecurity monitoring. The software is modified to eliminate the vulnerability. No other changes are made. There is no further impact to the software or device. The cybersecurity vulnerability is eliminated without changing the device’s functionality. Risk analyses determine there is no negative impact to the software or device.

Not significant

Software – cybersecurity

A connection attempt limit is added to an implanted cardiac pacing device that locks access to the device following a set number of failed connection attempts. This is done to prevent unauthorized access. The change is made to strengthen the cybersecurity of the device. However, it may restrict legitimate connections to the device that are clinically important. This change also impacts the device’s safety risk assessment.

Significant

Changes in materials for non in vitro diagnostic devices

A change in material type or formulation may affect the chemical, physical and/or electromagnetic properties of the material. A material change may also influence the device’s:

Changes related to materials may be intentional or unintentional. Changes in a manufacturing process, equipment or material storage conditions may also affect the properties of device materials.

Examples:

All changes should be assessed separately.

The impact of a material change on the safety and/or effectiveness of the device also depends on:

Note: Please refer to other sections of this guidance document for a material change related to changes in:

The following flowchart and a detailed explanation outline the assessment you must undertake to determine if your change is significant.

Figure 1: Flowchart outlining the decision-making process for changes in material for non-IVDD devices

This figure shows where the device ID and the device identifiers, catalogue numbers or model numbers are located on an issued medical device licence.
Text description

Figure 1 is a flowchart containing 7 boxes with yes or no questions. Each answer advances the reader through the decision-making process.

Box 1 in the flowchart shows that a change to the sourcing or processing of materials of human or animal origin is significant. For more information, consult:

Box 2 concerns changes to the vendor or supplier. This type of change is not significant if it does not alter the material type, formulation, chemical composition or processing.

Starting from Box 3, the considerations relate to changes that altered the material type, formulation, chemical composition or processing.

First, as indicated in Box 3, you need to determine if the altered material will be in direct or indirect contact with body tissues or fluids.

However, in terms of device safety, there is no difference between direct and indirect contact with body tissues or fluids. A change that involves either type of contact could be significant, pending further considerations.

The change is not significant if the altered material does not have direct or indirect contact with body tissues or fluids and does not cause other changes that could be significant.

If the altered material is in contact with body tissues or fluids, then Box 4 asks if the altered material is used in an authorized device with a similar design for the same indications by the same manufacturer. The change is not significant if the 2 devices are manufactured by the same process and the material is used in the same physical configuration for the same indication in the 2 devices. In addition, there must have been no concerns about the biocompatibility of the authorized device.

The authorized device must have all of the following:

Note: A material change may affect how it interacts with adjacent materials that bond together. Introducing a material used previously in an authorized device to a device where the interactions with adjacent materials are different is a significant change.

If the device doesn't meet those conditions, then Box 5 asks if the altered material is used in an invasive device that is to:

If a device meets these conditions, it is a significant change, unless:

If the device doesn't meet those conditions, then Box 6 asks about biocompatibility concerns. A material change associated with a new or increased biocompatibility concern is a significant change:

For example, if an altered material introduces a new leachable that needs to be tested:

Finally, according to Box 7, if a material change results in other changes such as device specifications, performance, manufacturing process or labelling, consult other sections of this guidance where appropriate.

Table of examples

Device  Proposed change Significant or not significant

Hip implant

Change to the supplier of titanium alloy used to manufacture the hip stem. The material continues to conform with ASTM F136. 

Not significant

Peripherally inserted central catheter (PICC)

Colourant change to the insertion hub of a PICC that is part of the fluid path for fluid administration or withdrawal from a patient. The colourant complies with the FDA List 7 Color additives exempt from certification (unless otherwise indicated) and permanently listed for use in medical devices.

Not significant

Peripherally inserted central catheter (PICC) 

Colourant change to the flush port of a PICC. The flush port is an access port for flush syringes for IV line clearance or volume block and is not to be used for fluid administration or withdrawal from a patient. The colourant does not come in contact with body tissues or fluids directly or indirectly.

Not significant

Cardiovascular 
catheter 

Polyether block amide (PEBA) changed to polyether ether ketone (PEEK) in a component in a cardiovascular catheter that comes in contact with the cardiovascular system. The same material has been used in another authorized cardiovascular catheter by the same manufacturer.

Significant

Dental composite

Change to the concentration of a dental composite component. The device performance specifications are the same. The device stays in the body for more than 30 days.

Significant

Dermal filler 

Change to the crosslinking degree of hyaluronic acid used in a dermal filler. The formulation of the device is the same. The device is absorbed by and stays in the body for more than 30 days.

Significant

Breast implant 

Change to a processing equipment, which alters the specifications of the material used in a breast implant. The device stays in the body for more than 30 days.

Significant

Hemodialysis system/console

Change to the material used in the pressure sensor of a hemodialysis system fluid path. The altered material is identical to material used in a temperature sensor in another authorized hemodialysis system fluid path. Both hemodialysis systems are made by the same manufacturer with the same indications for use. They only differ in software features. The material does not contact body fluids for more than 30 days. There is no other concomitant change.

Not significant

Oxygenator

Change to the membrane material used in the gas-exchange part of an oxygenator. The material has not been used in another authorized oxygenator. It is not in contact with body fluids for more than 30 days. However, there is an increased biocompatibility concern about this material.

Significant

Suture needle

Material used in a suture needle changed from 304 Stainless Steel to 316 Stainless Steel. The device is not absorbed by and does not stay in the body for more than 30 days. The manufacturer has another needle authorized that uses the same 316 Stainless Steel. The needle mechanical performance specifications are the same.

Not significant

Percutaneous transluminal angioplasty (PTA) balloon dilatation catheter

Ink used on the proximal marker of a PTA balloon dilatation catheter changed from white hot stamp ink to a black TP300 N50 ink. This ink is in contact with patients but has been used in a authorized similar PTA device by the same manufacturer. The device continues to meet finished product design specifications. The device is not absorbed by and does not stay in the body for more than 30 days.

Not significant

PTA balloon dilatation catheter

Material used in the hypotube of a PTA balloon dilatation catheter changed from 304L Stainless Steel to 304 Stainless Steel. The 304 Stainless Steel hypotube has been used in another authorized PTA catheter by the same manufacturer for the same indications. The device continues to meet the device design specifications. The device is not absorbed by and does not stay in the body for more than 30 days.

Not significant

Cardiovascular stent delivery system

Change to the material of the handle of a cardiovascular stent. There is no change to the design of the handle and the handle does not come in contact with the patient.

Not significant

Changes in materials for in vitro diagnostic devices

There is a distinction between in vitro diagnostic devices (IVDDs) and other devices regarding material changes.

Material changes of an IVDD include those made to:

Examples of material changes that are likely significant are changes:

Changes to the materials of an IVDD that cause a change to the design, manufacturing process, equipment, control procedures and/or labelling, including stability, are addressed in other sections of this guidance.

Table of examples

Device  Proposed change Significant or not significant

IVDD test kit

Change in the preservative of control material for which internal verification testing showed no change in performance characteristics, including stability. Also, no new hazards have been identified.

Not significant

IVDD test kit

Sodium azide added as a preservative to a reagent of the kit. This introduces a new hazard and requires that a warning be added to the labelling.

Significant

IVDD test kit

Change in PBS buffer within a test kit due to a supplier change with no change to the manufacturer’s specifications. Risk assessment and verification do not identify an impact on safety or effectiveness of the device and no new stability studies are warranted.

Not significant

IVDD test kit

Glass reagent bottles replaced with plastic reagent bottles to address breakage issues, resulting in manufacturer specifications changing for the container. A new stability study is required to establish stability of the reagent using the new container material.

Significant

IVDD test kit

Sample preparation changed to include a stabilizer. A risk assessment shows that adding a stabilizer can impact the assay’s performance.

Significant

IVDD test kit

Change to the sample extraction buffer pH is made to improve extraction efficiency. Risk assessment did not identify an impact on safety and verification activities did not identify an impact on effectiveness. No changes to labelling are made.

Not significant

IVDD test kit

External positive control source material changed from human antibody to non-human antibody. The change means a new stability study as the stability of the newly formulated control has not been established.

Significant

IVDD test kit

Primary packaging changed from a less water-permeable to a more water-permeable material (for example, highly impermeable foil to plastic). Risk assessment and verification activities identify an impact on device stability.

Significant

Changes to labelling

Labelling includes:

Labelling changes are often triggered by:

You must consider each labelling change separately, as well as the total impact of the labelling changes over time.

Changes that trigger a labelling change should also be assessed against all other relevant sections of this guidance.

Changes concerning indications for use, intended use and clinical benefits

Changes to the intended use or indications for use are significant unless they are within the current scope of indications set out in the labelling. This follows the general principles where:

A significant change would be to add a new subgroup of patients to the indications or intended use. This group may be considered higher risk than the more general population group identified in the previous indications or intended use.

However, this change is likely not significant if the identified subgroup of patients was added for clarification and would have been covered within the indications for use for the authorized device.

Example of a significant change:

Example of a non-significant change:

In general, reducing indications for use to a subset of the original population is a non-significant change.

Despite this, the following would be considered a significant change:

Examples of other significant changes to labelling:

Examples of changes to warnings and precautions that may not be significant:

Examples of changes to warnings and precautions that are significant:

Adding a clinical benefit to labelling may introduce new claims or expand existing claims made in the indications for use. This is likely to be a significant change in most cases.

Examples of a significant change:

Example of a non-significant change:

New labelling may be introduced that targets different populations (such as patient-specific labelling) or is created for marketing purposes.

Examples of a non-significant change:

For IVDDs, the following are examples of changes that are likely significant:

Changes concerning reprocessing, sterilizing, cleaning or disinfection

Medical devices that are reprocessed, sterilized, cleaned or disinfected by the end user must include manufacturer validated guidance on the cleaning, disinfection and/or sterilization process.

Consult:

Changes to the recommended sterilizing, cleaning or disinfecting method or product as previously submitted and reviewed in relation to an authorized class III or IV device may be deemed significant. Removing a recommended method or product from the labelling or package insert may not be a significant change if alternate, validated options authorized in Canada are still included in the labelling.

Changes concerning other regulatory jurisdictions

For information on adding a clinical benefit to labelling, please refer to the subsection on Changes concerning indications for use, intended use and clinical benefits. Including additional languages required in other regulatory jurisdictions is not a significant change.

Changes related to references in the labelling are significant if they:

When it comes to references, the following are not significant changes:

You should also remove references to obsolete devices from the instructions for use once a device is no longer available for sale and use in Canada. This is not considered a significant change.

Changes concerning the useful life of a product

The following are examples of a significant change:

Examples of a non-significant change:

Changes concerning compatible devices

You should clearly communicate within the labelling what devices are available in Canada. Information should clearly indicate how the device can be used safely and effectively with authorized compatible devices that are available in Canada. Be sure to remove references to obsolete devices from the instructions for use once a device is no longer available for sale and use in Canada (not a significant change).

Adding references in the instructions for use to unauthorized devices or to devices that were previously authorized but are no longer available in Canada due to obsolescence are not significant changes as long as:

Adding compatibility claims to labelling of a Class III or IV device with a separately authorized Class III or IV device is often a significant change. This applies to both of the following situations:

Adding a compatibility claim with an authorized device that’s similar to a currently listed compatible device is not necessarily a significant change.

Adding a compatibility claim would not be significant if:

However, adding a new compatible device is likely significant if:

If you made a compatibility claim with a device from a different manufacturer, you should monitor for possible changes that would render the compatibility invalid. A significant change made by a third-party manufacturer to a compatible device could impact the safety and effectiveness of your device. This would be a significant change.

For changes in compatibility claims associated with a compatible Class II device, refer to the following section on Class III or IV amendments concerning separate compatible Class II devices.

Removing an approved compatibility claim may be considered a significant change if:

Changes concerning magnetic resonance

In general, a change to the magnetic resonance (MR) safety claim of an authorized medical device, in terms of scan conditions under which a scan may be safely performed, is significant.

A non-significant change would be where:

Table of examples

Device Proposed change Significant or not significant

All devices

Deleting a contraindication (such as "not for pediatric use"), deleting a contraindication against lip augmentation for a dermal filler or removing a contraindication against the use of a dental implant in patients who smoke.

Significant

All devices

Labelling change to include additional languages, other than French or English, required in other regulatory jurisdictions.

Not significant

Dermal filler

Deleting potential adverse events, such as granuloma formation.

Significant

Percutaneous aortic valve

Introducing an additional warning that the device could embolize if not deployed completely and confirmed under fluoroscopy.

Not significant

Stent graft

Modifying the indications for use to exclude femoral implantation, but this was previously indicated. This change is implemented due to safety concerns identified in post-market use.

Significant

Radiofrequency generator

The radiofrequency generator is approved for use with authorized radiofrequency probes for the indication of creating radiofrequency lesions in nervous tissue. Adding another mode to the generator to be used with other authorized radiofrequency probes that are approved for use in the intervertebral disc to coagulate and decompress disc material.

Significant

Radiofrequency probe

The radiofrequency probe is indicated for ablating nervous tissue (used peripherally). Modifying indications for use so that the probe may now be used in the central nervous system (for example, brain).

Significant

Structural heart defect device (PFO closure)

Modifying indications for use to clarify that it is to be used in adults older than 21 years of age. This is the patient population that would normally be considered within the scope of a general indication for use.

Not significant

Ventricular assist device

Modifying indications for use from an adult population to a population with a mass greater than 20 kg.

Significant

Total knee replacement

Adding clinical benefit to state that a total knee replacement has 80% reliability at 15 years.

Significant

Class III and IV amendments for compatible Class II devices

“Compatibility” is the ability of a device, when used with 1 or more other devices, to achieve the overall clinical purpose without the user having to modify or adapt any part of the combined devices.

An important requirement in demonstrating safety and effectiveness of medical devices that are to be used together is compliance with section 18 of the Medical Devices Regulations. Under section 18, a medical device intended to be used together with other medical devices must:

This requirement applies to compatible devices that reside on the same authorization and those that are authorized separately.

Refer to the following notice for the situation when the devices are authorized separately:

When a change is made to 1 or more of the compatible devices, you must ensure that the combination of medical devices continues to meet section 18 requirements. As well, as per item (d) of the definition of “significant change”, a change to the intended use of the device includes new or extended use.

This section of this guidance applies to the following 2 scenarios:

  1. the labelling of a Class III or Class IV device is being changed to indicate compatibility with a Class II device, even if the formal “intended use and/or indications for use” statement stays the same
  2. a change is being made to an authorized Class II device that is already indicated as compatible with an authorized Class III or IV device

To evaluate whether either of these 2 scenarios are significant changes for the Class III or IV authorization, you must assess if they may affect the safety or effectiveness of the Class III or IV device. Consider the following factors:

You should identify and analyze how such differences could impact safety and effectiveness of the overall system. This includes assessing whether the change may affect such factors as compatibility, performance and risk mitigations.

Table of examples

Device Proposed change Significant or not significant

Class II swab used to collect endocervical specimens

The validation of the swab for use with a specific Class III Chlamydia test was reviewed as part of the application for that test. Change made to the specifications of the swab (for example, in material or design) that could impact the safety and/or performance of the test.

Significant

Class II administration set

Adding a new Class II administration set to a Class III programmable infusion pump’s list of compatible administration sets. Following an engineering analysis and risk assessment, the manufacturer determined that the design parameters of the new administration set fall within the design parameters of the administration sets. These are already indicated as compatible with the pump (for example, tubing dimensions and material, number and characteristics of integrated components, manufacturing processes, including sterilization). Verification testing was done to verify that the pump performs as expected with the new administration set using the same testing activities that were reviewed in a previous application. The testing did not produce any unexpected results. There are no changes to the intended use or indications for use compared to the authorized administration sets.

Not significant

Class II insulin infusion set

Adding a new Class II insulin infusion set to a Class III or IV programmable insulin pump’s list of compatible infusion sets. The new infusion set is indicated for a longer wear period than the infusion sets that are previously indicated as compatible with the pump and longer than stated in the pump labelling. The longer wear period necessitated new testing activities to support the safety and effectiveness of the therapy delivered by the overall system.

Significant

Class II delivery catheter indicated for use in the delivery of transcatheter aortic valves

A Class IV transcatheter aortic valve is authorized as compatible with a Class II delivery catheter. In this example, the Class II device’s labelling specifies the authorized access routes, but the Class IV device’s labelling does not. Introducing a design change to the Class II delivery catheter system allows for a different access route than the one authorized in the initial application. The design changes to the delivery catheter impact the clinical performance of the devices when they are used together, including a potential increase in risk and complications in delivering the transcatheter aortic valve. New clinical testing was required to support the safety and effectiveness of the overall system.

Significant

Change in reprocessing of a Class III or IV device

An authorized Class III or IV device is labelled for use with a Class II disinfecting device. A change in the specifications of the Class II disinfecting device could impact the performance or safety of the Class III or IV device when it is disinfected using the Class II device.

Significant

Class II surgical stapler handle

Change to an authorized Class II surgical stapler handle, which could impact the performance of the compatible Class III stapler reloads.

Significant

Class II sound processor

Change to the software of a sound processor authorized separately as a Class II device, but compatible with a Class III cochlear implant and/or a Class IV brain stem implant, as well as Class II hearing aids. The change in software could impact the performance of the Class III or IV systems or safety mitigations.

Significant

Changes to diagnostic ultrasound systems

Health Canada recognizes there are a number of diagnostic ultrasound systems authorized in Canada that have an established safety profile in Canada and in other regulatory jurisdictions. Consequently, a change made to a diagnostic ultrasound system is not significant if all of the following apply:

Table of examples

Device Proposed change Significant or not significant

Diagnostic ultrasound system

Adding M-mode imaging to the modes of operation of the device.

Not significant

Diagnostic ultrasound system

Adding a new transducer with the same indications for use as the system’s. The acoustic output and interrogation parameters are within the defined limits.

Not significant

Diagnostic ultrasound system

Adding a software feature that uses shear-wave elastography to measure liver stiffness.

Significant

Page details

Date modified: